Spontaneous Late Dissociation of the Tibial Insert After High-Flex Posterior-Stabilized Genesis II Total Knee Arthroplasty
- Lee, Dae-Hee;
- Lee, Tze Gin;
- Park, Sung-Joon;
- Han, Seung-Beom
Citations
WEB OF SCIENCE
7Citations
SCOPUS
9초록
We report a rare complication of spontaneous late dissociation of the tibial insert 2 years after total knee arthroplasty using a high-flex posterior-stabilized Genesis II prosthesis (Smith & Nephew, Memphis, Tenn). It appears that 2 factors may have contributed to dissociation, namely, incomplete seating of the insert and the design of the prosthesis, which involves a shallow anterior tab snap-fit locking mechanism and thin dovetail lips. (C) 2013 Elsevier Inc. All rights reserved.