상세 보기
Preliminary results of a prototype C-shaped PET designed for an in-beam PET system
- Kim, Hyun-Il;
- Chung, Yong Hyun;
- Lee, Kisung;
- Kim, Kyeong Min;
- Kim, Yongkwon;
- 외 1명
WEB OF SCIENCE
3SCOPUS
4초록
Positron emission tomography (PET) can be utilized in particle beam therapy to verify the dose distribution of the target volume as well as the accuracy of the treatment. We present an in-beam PET scanner that can be integrated into a particle beam therapy system. The proposed PET scanner consisted of 14 detector modules arranged in a C-shape to avoid blockage of the particle beam line by the detector modules. Each detector module was composed of a 9x9 array of 4.0 mm x 4.0 mm x 20.0 mm LYSO crystals optically coupled to four 29-mm-diameter PMTs using the photomultiplier-quadrant-sharing (PQS) technique. In this study, a Geant4 Application for Tomographic Emission (GATE) simulation study was conducted to design a C-shaped PET scanner and then experimental evaluation of the proposed design was performed. The spatial resolution and sensitivity were measured according to NEMA NU2-2007 standards and were 6.1 mm and 5.61 cps/kBq, respectively, which is in good agreement with our simulation, with an error rate of 12.0%. Taken together, our results demonstrate the feasibility of the proposed C-shaped in-beam PET system, which we expect will be useful for measuring dose distribution in particle therapy. (C) 2016 Elsevier B.V. All rights reserved.
키워드
- 제목
- Preliminary results of a prototype C-shaped PET designed for an in-beam PET system
- 저자
- Kim, Hyun-Il; Chung, Yong Hyun; Lee, Kisung; Kim, Kyeong Min; Kim, Yongkwon; Joung, Jinhun
- 발행일
- 2016-06-21
- 유형
- Article
- 권
- 822
- 페이지
- 57 ~ 62