Effect of baseplate size on primary glenoid stability and impingement-free range of motion in reverse shoulder arthroplasty

  • Chae, Soo-Won; 
  • Kim, Soung-Yon; 
  • Lee, Haea; 
  • Yon, Joung-Ro; 
  • Lee, Juneyoung; 
  • 외 1명
Citations

WEB OF SCIENCE

28
Citations

SCOPUS

41

초록

Background: Use of a baseplate with a smaller diameter in reverse shoulder arthroplasty is increasing, especially in patients with a small glenoid or glenoid wear. However, the effect of a smaller baseplate on stability of the glenoid component has not been evaluated. Thus, the purpose of this study was to determine whether a smaller baseplate (25 mm) is beneficial to the initial stability of the glenoid component compared to that with a baseplate of a commonly used size (29 mm). Methods: Micromotion of glenoid components attached to 14 scapulae of fresh-frozen cadavers was measured and compared between 25- and 29-mm baseplates in biomechanical testing. Impingement-free range of motion in abduction, adduction, internal rotation, and external rotation was evaluated by using a simulated computer model constructed based on the same fresh-frozen cadavers used in biomechanical testing. Results: Micromotion at the inferior third of the glenoid-glenosphere interface was higher in the 29-mm baseplate group than in the 25-mm baseplate group during both 0.7- and 1-body weight cyclic loading in biomechanical testing. Adduction deficit was smaller, and total impingement-free range of motion from abduction to adduction and rotation were greater in the 25-mm baseplate group than in the 29-mm baseplate group in the simulated computer model. Conclusions: Use of a baseplate with a smaller diameter (25 mm) in reverse shoulder arthroplasty is suitable for improving the primary stability of the glenoid component. With a smaller baseplate, impingement-free range of motion is optimized in a smaller glenoid.

키워드

Reverse shoulder arthroplasty; Smaller baseplate; Biomechanical testing; Simulated computer model; BIOMECHANICAL EVALUATION; COMPONENT FIXATION; IN-VITRO; PROSTHESIS; DESIGN; POSITION; FORCES; SCREW; ABDUCTION; ARTHRITIS
제목
Effect of baseplate size on primary glenoid stability and impingement-free range of motion in reverse shoulder arthroplasty
저자
Chae, Soo-Won; Kim, Soung-Yon; Lee, Haea; Yon, Joung-Ro; Lee, Juneyoung; Han, Seung-Ho
DOI
10.1186/1471-2474-15-417
발행일
2014-12-09
유형
Article
저널명
BMC Musculoskeletal Disorders
권
15