EVALUATION OF THE ACCURACY OF DOSE DELIVERY FOR IMRT BASED ON TRANSIT DOSIMETRY

  • Baek, Tae Seong; 
  • Chung, Eun Ji; 
  • Koh, Eui Kwan; 
  • Seo, Jungju; 
  • Yoon, Myonggeun
Citations

WEB OF SCIENCE

2
Citations

SCOPUS

2

초록

The authors have developed and evaluated a new method that uses transit dose in the treatment room in verification of dose delivery to patients. Five intensity modulated radiotherapy (IMRT) plans were selected from actual cancer patients, and transit dose was measured using MapCHECK 2 and an anthropomorphic phantom. The criteria used as a verification tool for the dose delivery to the patient were gamma-index-based dose comparison between the computed dose and measured dose. When the doses were delivered to an anthropomorphic phantom normally, the average passing rate was 95.2% based on a gamma index analysis. This feasibility study suggested that transit dose-based quality assurance can provide information about the accuracy of an inhomogeneity correction algorithm and patient positioning during treatment, allowing its use as a verification tool for actual dose delivery to patients in the treatment room.

키워드

exposure, radiation; physics, medical; quality assurance; radiation, medical; MODULATED RADIATION-THERAPY; QUANTITATIVE-EVALUATION; VOLUME RESTRICTIONS; QUALITY-ASSURANCE; BEAM WEIGHTS; OPTIMIZATION; VERIFICATION; DISTRIBUTIONS; TUMOR; FILM
제목
EVALUATION OF THE ACCURACY OF DOSE DELIVERY FOR IMRT BASED ON TRANSIT DOSIMETRY
저자
Baek, Tae Seong; Chung, Eun Ji; Koh, Eui Kwan; Seo, Jungju; Yoon, Myonggeun
DOI
10.1097/HP.0000000000000098
발행일
2014-09
유형
Article
저널명
Health Physics
권
107
호
3
페이지
200 ~ 205