Sensing of Airborne Nanoparticles Using Miniaturized Whipple Double Condenser

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

We have demonstrated the sensing of airborne nanoparticles using a miniaturized Whipple double condenser ( MWDC). Each condenser ( 17 mm long, 2.2 mm wide, and 1 mm tall) was fabricated using photolithography and copper electroplating. The MWDC was theoretically shown to function within the laminar flow regime and capable of detecting 10- and 20-nm particles at a flow rate of 0.3 LPM. Monodispersed isopropyl alcohol and oleic acid airborne nanoparticles ( 10 and 20 nm) were generated using a TSI Model 9302 Single Jet Atomizer and TSI 3080 Electrostatic Classifier, and were electrically charged with Kr-85 isotope. The calculated and experimental particle pass through ratio ( PPR) agreed reasonably well. Using a Keithley 616 electrometer, similar threshold condenser voltage ( maximum second condenser current) was observed in both experimental PPR as well as the second condenser current versus condenser voltage plot. The MWDC also displayed characteristic Whipple curves in its experimental I/V versus kc plots. These data demonstrate the feasibility of the microfabricated double condenser design as the basis of a miniaturized nanoparticle sizing and counting device.

키워드

Whipple double condenserminiaturizedairborne nanoparticlesparticle pass through ratioULTRAFINE PARTICLESAEROSOLHEALTHSIZERAIR
제목
Sensing of Airborne Nanoparticles Using Miniaturized Whipple Double Condenser
저자
Barrett, TerenceChua, BeeleeHolmen, Britt A.
DOI
10.1109/JSEN.2016.2584580
발행일
2016-09-15
유형
Article
저널명
IEEE Sensors Journal
16
18
페이지
6990 ~ 6996