USE OF CLINICAL EXPOSURE INDEX AND DEVIATION INDEX BASED ON NATIONAL DIAGNOSTIC REFERENCE LEVEL AS DOSE-OPTIMIZATION TOOLS FOR GENERAL RADIOGRAPHY IN KOREA

  • Park, Hyemin
  • Yoon, Yongsu
  • Kim, Jungmin
  • Kim, Jungsu
  • Jeong, Hoiwoun
  • 외 2명
Citations

WEB OF SCIENCE

7
Citations

SCOPUS

9

초록

The International Electrotechnical Commission introduced the concepts of exposure index (EI), target exposure index (EIT) and deviation index (DI) to manage and optimize patient dose in real time. In this study, we have proposed an appropriate method for setting the EIT based on the Korean national diagnostic reference levels (DRLs). Furthermore, we evaluated the use of clinical EI, EIT and DI as tools for patient dose optimization in clinical environments by observing the changes in DI with those in EIT. According to the Korean national exposure conditions, we conducted experiments on three representative radiographic examinations (chest posterior-anterior, lateral and abdomen anterior-posterior) of clinical environments. As the exposure conditions and DRLs varied, the clinical EI, EIT and DI also varied. These results reveal that the clinical EI, EIT and DI can be used as tools for optimizing the patient dose if EIT is periodically and properly updated.

키워드

DIGITAL RADIOGRAPHY
제목
USE OF CLINICAL EXPOSURE INDEX AND DEVIATION INDEX BASED ON NATIONAL DIAGNOSTIC REFERENCE LEVEL AS DOSE-OPTIMIZATION TOOLS FOR GENERAL RADIOGRAPHY IN KOREA
저자
Park, HyeminYoon, YongsuKim, JungminKim, JungsuJeong, HoiwounTanaka, NobukazuMorishita, Junji
DOI
10.1093/rpd/ncaa185
발행일
2020-10
유형
Article
저널명
Radiation Protection Dosimetry
191
4
페이지
439 ~ 451