방사선치료의 다양한 EPID 영상 질평가

Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy
  • 손순룡
  • 최관우
  • 김정민
  • 정회원
  • 권경태
  • 외 7명

초록

In megavoltage (MV) radiotherapy, delivering the dose to the target volume is important while protecting the surrounding normal tissue. The purpose of this study was to evaluate the modulation transfer function (MTF), the noise power spectrum (NPS), and the detective quantum efficiency (DQE) using an edge block in mega-voltage X-ray imaging (MVI).We used an edge block, which consists of tungsten with dimensions of 19 (thickness) × 10 (length) × 1 (width) cm3 and measured the pre-sampling MTF at 6 MV energy. Various radiation therapy (RT) devices such as TrueBeamTM (Varian), BEAMVIEWPLUS (Siemens), iViewGT (Elekta) and Clinac®iX (Varian) were used. As for MTF results, TrueBeamTM(Varian) flattening filter free(FFF) showed the highest values of 0.46 mm-1and1.40mm-1for MTF 0.5 and 0.1. In NPS, iViewGT (Elekta) showed the lowest noise distribution. In DQE, iViewGT (Elekta) showed the best efficiency at a peak DQE and 1 mm-1DQE of 0.0026 and 0.00014, respectively.This study could be used not only for traditional QA imaging but also for quantitative MTF, NPS, and DQE measurement for development of an electronic portal imaging device (EPID).

키워드

변조전달함수잡음전력스펙트럼양자검출효율전자포털영상장치Modulation transfer function (MTF)Noise power spectrum (NPS)Detective quantum efficiency (DQE)Electronic portal imaging device (EPID)
제목
방사선치료의 다양한 EPID 영상 질평가
제목 (타언어)
Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy
저자
손순룡최관우김정민정회원권경태조정희이제희정재용김기원이영아손진현민정환
DOI
10.17946/JRST.2015.38.4.16
발행일
2015
저널명
방사선기술과학
38
4
페이지
451 ~ 461