Acoustic mixing in a dome-shaped chamber-based SAW (DC-SAW) device

Citations

WEB OF SCIENCE

33
Citations

SCOPUS

35

초록

The use of an open droplet system for surface acoustic wave (SAW)-based applications has been limited by droplet instability at high input power. This study introduces a dome-shaped chamber-based SAW (DC-SAW) device for the first time, which can be fabricated simply using a single adhesive tape and a drop of ultraviolet-curable material without soft lithography processes. The dome-shaped chamber device with a contact angle of 68 degrees enables the maximizing of the effect of SAW transmitted at a refraction angle of roughly 22 degrees, negating the droplet instability. The DC-SAW device was applied to acoustic mixing to estimate its capability. Acoustic mixing of two different fluids (i.e., deionized water and fluorescent particle suspension) was demonstrated in the dome-shaped chamber device. Moreover, the effect of flow rate and applied voltage on mixing performance was estimated. With the decreasing flow rate and increasing applied voltage, mixing performance was enhanced. At an applied voltage of 20 V, mixing indices were higher than 0.9 at a total flow rate of 300 mu l min(-1).

키워드

ACOUSTOFLUIDIC MICROMIXERMICROFLUIDIC CHANNELSESSILE DROPLETPARTICLEWAVES
제목
Acoustic mixing in a dome-shaped chamber-based SAW (DC-SAW) device
저자
Lim, HyunjungBack, Seung MinChoi, HyukNam, Jeonghun
DOI
10.1039/c9lc00820a
발행일
2020-01-07
유형
Article
저널명
Lab on a Chip
20
1
페이지
120 ~ 125