A flexible glucose sensor is fabricated using O-2 plasma-functionalized multiwalled carbon nanotube (MWCNT) films on polydimethylsiloxane (PDMS) substrates and its performance is electrochemically characterized. After enzyme immobilization, the GOD/MWCNT/Au/PDMS electrode exhibits a sensitivity of 18.15 mu A mm(-2)mM(-1) and a detection limit of 0.01 mM (signal to noise ratio was about 3). This high sensitivity may be attributed to a large enzyme loading and a higher electrocatalytic activity and electron transfer exhibited by O-2 plasma-functionalized CNTs than the pristine CNT, due to some oxygen-contained groups present on the O-2 plasma-functionalized CNT surface, which has been verified by XPS spectrum. (C) 2009 Elsevier B.V. All rights reserved.
키워드
- 제목
- Flexible electrochemical biosensors based on O-2 plasma functionalized MWCNT
- 저자
- Lee, Jun-Yong; Park, Eun-Jin; Lee, Cheol-Jin; Kim, Soo-Won; Pak, James Jungho; Min, Nam Ki
- 발행일
- 2009-05-29
- 유형
- Article; Proceedings Paper
- 저널명
- Thin Solid Films
- 권
- 517
- 호
- 14
- 페이지
- 3883 ~ 3887