Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B

Citations

WEB OF SCIENCE

9
Citations

SCOPUS

10

초록

We evaluated the performance of the STANDARD Q COVID/FLU Ag Combo test (Q Ag combo test) for the detection of SARS-CoV-2, influenza A, and influenza B using a single point-of-care device compared with real-time PCR. A total of 408 individuals, 55 positives with SARS-CoV-2, 90 with influenza A, 68 with influenza B, and 195 negatives for all viruses, participated. The Q Ag combo test demonstrated a high level of sensitivity of 92.73% and a specificity of 99.49% for the detection of SARS-CoV-2. When the number of days from symptom onset (DSO) was restricted to 0 < DSO <= 6, the sensitivity of the Q Ag combo test to detect SARS-CoV-2 was 100%, and when the Ct value of RdRp was <= 20, the sensitivity to detect SARS-CoV-2 was 93.10%. The Q Ag combo test results also demonstrated a sensitivity of 92.22% and a specificity of 100% for influenza A, a sensitivity of 91.18%, and a specificity of 99.49% for influenza B. The agreement analysis of the Q Ag combo test with the RT-PCR results demonstrated excellent outcomes, making it useful and efficient for the detection of SARS-CoV-2, influenza A, and influenza B.

키워드

rapid antigen testSARS-CoV-2influenza Ainfluenza Bpoint-of-careDIAGNOSISPCR
제목
Performance Evaluation of STANDARD Q COVID/FLU Ag Combo for Detection of SARS-CoV-2 and Influenza A/B
저자
Widyasari, KristinKim, SoominKim, SunjooLim, Chae Seung
DOI
10.3390/diagnostics13010032
발행일
2023-01-01
유형
Article
저널명
Diagnostics
13
1