상세 보기
Dose area product measurement for diagnostic reference levels and analysis of patient dose in dental radiography
- Han, Suchul;
- Lee, Boram;
- Shin, Gwisoon;
- Choi, Jonghak;
- Kim, Jungmin;
- ... Lee, Kisung;
- 외 3명
WEB OF SCIENCE
20SCOPUS
24초록
In this study, diagnostic reference levels (DRLs) were suggested and patient doses were analysed through the dosearea product value in dental radiography. In intraoral radiography, at three sites, i.e. molar, premolar and incisor on the maxilla and acquired third quartile values: 55.5, 46 and 36.5 mGy cm(2), respectively, were measured. In panoramic, cephalometric and cone beam computed tomography, the values were 120.3, 146 and 3203 mGy cm(2) (16 18 cm), respectively. It has been shown that, in intraoral radiography, the patient dose changes proportionally to the value of mA s, but the change in extraoral radiography in response to mA s could not be confirmed. The authors could confirm, however, the difference in dose according to the manufacturer in all dental radiography examinations, except for panoramic radiography. Depending on the size of hospital, there were some differences in patient dose in intraoral radiography, but no difference in patient dose in extraoral radiography.
키워드
- 제목
- Dose area product measurement for diagnostic reference levels and analysis of patient dose in dental radiography
- 저자
- Han, Suchul; Lee, Boram; Shin, Gwisoon; Choi, Jonghak; Kim, Jungmin; Park, Changseo; Park, Hyok; Lee, Kisung; Kim, Youhyun
- 발행일
- 2012-07
- 유형
- Article
- 권
- 150
- 호
- 4
- 페이지
- 523 ~ 531