Development of carbon nanoparticles-based soluble solid-phase immune sensor for the quantitative diagnosis of inflammation

  • Kim, S.
  • Koo, M.
  • Tak, Y.
  • Jang, S.
  • Park, J.
  • ... Hwang, K.Y.
  • 외 1명
Citations

WEB OF SCIENCE

5
Citations

SCOPUS

5

초록

Quantitative immunodiagnosis is one of the most commonly used methods for in vitro diagnostics. Various bioanalytical methods have been developed to quantitatively diagnose immune analytes; however, they require blood dilution pretreatment, reaction mixing, complicated experimental steps, and can cause diagnostic errors due to the hook effect. To address this issue, we introduced a simple immunoassay based on carbon nanoparticles (CNPs). The assay was designed to have high flexibility for use in various in vitro diagnostic devices by constructing a soluble solid-phase immune sensor with high solubility using antibody-conjugated CNPs and polymer materials. Excellent performance was achieved using a free-antibody system with dual calibration. To verify the performance of this method with high reliability, canine C-reactive protein was selected as the immune analyte. Interestingly, our method efficiently mitigated the hook effect with outstanding performance in a one-step reaction without blood dilution or reaction mixing. The detection range of the target can be effectively controlled using free antibodies. Therefore, our CNP-based immunodiagnosis method may advance the commercialization of point-of-care immune biosensors. © 2022 The Authors

키워드

BiosensorC-reactive proteinCarbon nanoparticlesHook effectIn vitro diagnosticsPoint-of-Care
제목
Development of carbon nanoparticles-based soluble solid-phase immune sensor for the quantitative diagnosis of inflammation
저자
Kim, S.Koo, M.Tak, Y.Jang, S.Park, J.Hwang, K.Y.Park, S.
DOI
10.1016/j.bios.2022.114975
발행일
2023-02-15
유형
Article
저널명
Biosensors and Bioelectronics
222