Analysis of the Dosimetric Characteristics of Normoxic Polymer Gel by Magnetic Resonance Images

Citations

WEB OF SCIENCE

3
Citations

SCOPUS

3

초록

A new type of polymer gel that eliminates the oxygen chemically instead of using complicated facilities was recently introduced. This study utilized a polymer gel fabricated in a laboratory as a reliable dosimeter for measuring therapeutic radiation. Studies on the optimal MR scan conditions was conducted. The reproducibility, the accuracy, and the dose rate dependency of the normoxic polymer gel dosimeter were evaluated. The relative dose uncertainty and the relative dose resolution of the polymer gel were also evaluated by scanning it under optimal MR scan conditions. The R2 calculated by using echo time (TE) of 20, 40, 60 80, and 100 ms versus dose was linear for 12 similar to 20 Gy. The polymer gel composed in this study can be used as a reliable dosimeter if it is scanned under optimal MR, scan conditions.

키워드

Normoxic polymer gelThree-dimensional dosimetryDosimetric characteristicsMagnetic resonance imagesRadiotherapyRADIATION-DOSE DISTRIBUTIONSMRIRESOLUTIONOPTIMIZATION
제목
Analysis of the Dosimetric Characteristics of Normoxic Polymer Gel by Magnetic Resonance Images
저자
Cho, Sam JuShim, Su JungKim, Chul YongYang, Dae SikPark, Young JeLee, SukLee, Sang HoonLim, SangwookCho, Kwang HwanChoi, Jin HoLee, Rena
DOI
10.3938/jkps.56.874
발행일
2010-03
유형
Article
저널명
Journal of the Korean Physical Society
56
3
페이지
874 ~ 879