Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty A Retrieval Study Using 3-Dimensional Laser Scanning

  • Yun, Ho Hyun
  • Shon, Won Yong
  • Yoon, Jung Ro
  • Yang, Jae-Hyuk
  • Lim, Deuk-Soo
Citations

WEB OF SCIENCE

6
Citations

SCOPUS

9

초록

The purpose of this study was to evaluate the reliability of the Power Point (PP) (2007 Version; Microsoft, Redmond, Wash) method for measuring polyethylene liner wear after total hip arthroplasty. Seventeen retrieved polyethylene liners were included in this study. Wear volumes were calculated using the PP, the Dorr and Wan, and 3-dimensional (3D) laser scanning methods. Spearman correlation coefficients for wear volume results indicated strong correlations between the PP and 3D laser scanning methods (range, 0.89-0.93). On the other hand, Spearman correlation analysis revealed only moderate correlations between the Dorr and Wan and 3D laser scanning methods (range, 0.67-0.77). The PP method can be used to monitor linear wear after total hip arthroplasty and could serve as an alternative method when computerized methods are not available.

키워드

hiptotal hip arthroplastywearwear measurementPowerPoint methodCROSS-LINKED POLYETHYLENEBACKED ACETABULAR CUPSRADIOGRAPHIC MEASUREMENTFEMORAL-HEADVOLUMETRIC WEARPENETRATIONREPLACEMENTCOMPONENTSVALIDATIONACCURACY
제목
Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty A Retrieval Study Using 3-Dimensional Laser Scanning
저자
Yun, Ho HyunShon, Won YongYoon, Jung RoYang, Jae-HyukLim, Deuk-Soo
DOI
10.1016/j.arth.2011.12.025
발행일
2012-09
유형
Article
저널명
Journal of Arthroplasty
27
8
페이지
1530 ~ 1537