Ultrasonographic measurement of the cross-sectional area of the axillary nerve

  • Yoon, Joon Shik
  • Park, Jung Hyun
  • Levin, Joshua
  • Roh, Eugene Y.
  • Park, Hye Jung
  • 외 1명
Citations

WEB OF SCIENCE

5
Citations

SCOPUS

6

초록

Introduction The objectives of this study were to determine normal reference values for ultrasonographic measurement of the cross-sectional area (CSA) of the axillary nerve and to standardize the measurement methods. Methods Sixty healthy volunteers were evaluated. Ultrasonography was performed with the shoulder positioned in 100 degrees-120 degrees abduction and 90 degrees external rotation. The CSA of the axillary nerve was measured bilaterally. Results The normal CSA of the right axillary nerve was 2.9 +/- 1.1 mm(2). The side-to-side discrepancy was 22.8% +/- 17.8%. Discussion These reference values may be helpful for investigating pathologies involving the axillary nerve.

키워드

axillary nervenerve cross-sectional areareference valuesshoulderultrasonographyREFERENCE VALUESSURGICAL ANATOMYULNAR NERVEBRANCHESINJURIES
제목
Ultrasonographic measurement of the cross-sectional area of the axillary nerve
저자
Yoon, Joon ShikPark, Jung HyunLevin, JoshuaRoh, Eugene Y.Park, Hye JungWon, Sun Jae
DOI
10.1002/mus.26480
발행일
2019-07
유형
Article
저널명
Muscle and Nerve
60
1
페이지
95 ~ 98