Programmable High Voltage Distribution for Photodetectors in a 1 mm Resolution Clinical PET System

  • Lau, Frances W. Y.; 
  • Vandenbroucke, Arne; 
  • Yeom, Jung-Yeol; 
  • Reynolds, Paul D.; 
  • Hsu, David; 
  • 외 2명
Citations

WEB OF SCIENCE

5
Citations

SCOPUS

6

초록

We present a scalable, cost-effective, and compact high voltage (HV) distribution design for a system that requires many HV channels for photodetector biasing. A digital-to-analog converter (DAC) with a 100 V output range is floated with its reference at -1800 V, providing outputs between -1800 V and -1700 V. As a consequence, the DAC control and power signals are referenced to V. This design was implemented in our 1 mm resolution clinical PET imaging system with detectors comprised of LYSO scintillation crystals coupled to position-sensitive avalanche photodiodes (PSAPDs). The programmable bias voltage outputs have only 20 m V-pp of bias ripple. This design enables all PSAPDs to be biased at their optimal bias voltage, allowing us to achieve a standard deviation of only 2.3% in the energy resolution measured across all 512 PSAPDs in the subsystem.

키워드

Avalanche photodiode; high-voltage techniques; photodetectors; positron emission tomography (PET); THERMAL REGULATION; DESIGN; PERFORMANCE
제목
Programmable High Voltage Distribution for Photodetectors in a 1 mm Resolution Clinical PET System
저자
Lau, Frances W. Y.; Vandenbroucke, Arne; Yeom, Jung-Yeol; Reynolds, Paul D.; Hsu, David; Innes, Derek; Levin, Craig S.
DOI
10.1109/TNS.2015.2440437
발행일
2015-10
유형
Article
저널명
IEEE Transactions on Nuclear Science
권
62
호
5
페이지
1989 ~ 1994