Performance evaluation of three i-SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer

Citations

WEB OF SCIENCE

4
Citations

SCOPUS

4

초록

Background Capillary blood is the most commonly used sample for point-of-care (POC) glucometers. However, in critically ill patients, the glucose levels measured from capillary blood may not be reliable. Thus, we aimed to evaluate and compare the accuracy of glucose levels measured with POC glucometers and the YSI 2300 glucose analyzer using leftover arterial blood samples. Methods In total, 100 leftover heparinized arterial blood samples were used to evaluate the performance of three i-SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU-CHEK(R) Inform II glucometer. The reference value was obtained using the YSI 2300 glucose analyzer. The results were analyzed based on International Organization for Standardization 15197:2013 guidelines. Results More than 95% of results obtained using POC glucometers were within +/- 15 mg/dL of the reference value for glucose concentrations <100 mg/dL and within +/- 15% of the reference value for glucose concentrations >= 100 mg/dL. In the consensus error grid analysis, more than 99% of results were found to be within zones A and B. An excellent correlation was found between the values obtained using POC glucometers and the YSI 2300 glucose analyzer (R-2 > .99). Conclusion The i-SENS glucometers showed stable and accurate results when leftover arterial blood samples were used. Therefore, POC glucometers could be useful in critical care settings, such as intensive care units, where arterial samples are routinely used.

키워드

arterial blood i-SENSglucometerPOCTpoint-of-careMETERS
제목
Performance evaluation of three i-SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
저자
Kim, Ha NuiMoon, Kyung ChulYoon, Soo-Young
DOI
10.1002/jcla.23356
발행일
2020-08
유형
Article
저널명
Journal of Clinical Laboratory Analysis
34
8