Periodic reference subtraction method for efficient background fixed pattern noise removal in Fourier domain optical coherence tomography

Citations

WEB OF SCIENCE

10
Citations

SCOPUS

9

초록

We have demonstrated an efficient scheme of automatically removing fixed pattern noise in Fourier domain optical coherence tomography by using a periodic reference spectrum subtraction method. By periodically acquiring the reference spectra using a separate light absorber placed to the right of the scan lens, we were able to adaptively compensate the background fixed pattern due to the spectral intensity variation of the source. The adaptive removal of fixed pattern noise was effectively performed by controlling the reference spectrum acquisition rate (R). A seawater pearl was used for a test sample under an intentional abrupt source power change to validate the proposed method. Based on this method, it is possible to perform immediate cold start scanning because it is not necessary for a stabilization period of the light source, as well as a manual process of reference spectrum acquisition for obtaining clear image under unstable environment. (C) 2011 Elsevier B.V. All rights reserved.

키워드

Optical coherence tomographyFixed pattern noiseFD-OCTReference spectrum subtractionOPHTHALMOLOGYOCT
제목
Periodic reference subtraction method for efficient background fixed pattern noise removal in Fourier domain optical coherence tomography
저자
Kim, Ji-hyunHan, Jae-HoJeong, Jichai
DOI
10.1016/j.optcom.2011.12.045
발행일
2012-04-01
유형
Article
저널명
Optics Communications
285
7
페이지
2012 ~ 2016