NATIONAL DIAGNOSTIC REFERENCE LEVELS AND ACHIEVABLE DOSES FOR 13 ADULT CT PROTOCOLS AND A PAEDIATRIC HEAD CT PROTOCOL: NATIONAL SURVEY OF KOREAN HOSPITALS

  • Kim, Jung Su
  • Lee, Sang Kyung
  • Kim, Sun-Ki
  • Yoo, Seung-Min
  • Kim, Jung Min
  • 외 1명
Citations

WEB OF SCIENCE

11
Citations

SCOPUS

11

초록

To develop a second set of diagnostic reference levels (DRLs) and achievable doses (ADs) for 13 adult computed tomography (CT) protocols and a paediatric head CT protocol in Korea. A survey of 13,625 CT examinations was performed based on 13 adult CT protocols and a paediatric non-contrast brain CT protocol using 369 CT systems, with patients grouped according to age. Most CT protocols in this survey had DRLs similar to those reported in other countries. However, chest and abdomen-pelvic CT had lower DRLs than those reported in the first Korean national survey and those from other countries. Paediatric non-contrast brain CT in each age group, with the exception of the 11-15-year age group, had lower DRLs than those reported in other countries. The DRLs presented here are similar to (or lower than for some protocols) those reported in the first Korean national survey and those from other countries.

키워드

TOMOGRAPHY
제목
NATIONAL DIAGNOSTIC REFERENCE LEVELS AND ACHIEVABLE DOSES FOR 13 ADULT CT PROTOCOLS AND A PAEDIATRIC HEAD CT PROTOCOL: NATIONAL SURVEY OF KOREAN HOSPITALS
저자
Kim, Jung SuLee, Sang KyungKim, Sun-KiYoo, Seung-MinKim, Jung MinYoon, Sang-Wook
DOI
10.1093/rpd/ncz156
발행일
2019-12
유형
Article
저널명
Radiation Protection Dosimetry
187
2
페이지
220 ~ 229