세포 호흡량 측정용 다채널 측정 시스템 개발

Development of a multi channel measurement system for the cellular respiration measurement
  • 남현욱
  • 박정일
  • 김영미
  • 박정호

초록

This paper describes a multi channel measurement system which can measure the cellular respiration level in a solution containing cells by using a Clark-type sensor with the solution temperature control unit. The Clark-type sensor can measure the cellular respiration level in the solution because it can measure the reduction current depending on the dissolved oxygen level in the solution. This measurement system was maintained the temperature within ± 0.1 oC of the setting temperature value by on/off control method in order to measure the precise cellular respiration level. The measurement system showed that the applied voltage to the working electrode was very stable(−0.8 V± 0.0071 V) by using proportional control method. From the current measurement, the response time and the linearity correlation coefficient were 25 sec and 0.94, respectively, which are very close to the results of the commercial product. Using this system and the fabricated Clarktype sensor, the average ratio of the uncoupled OCR(oxygen consumption rate) to the coupled OCR was 1.35 and this is almost the same as that obtained from a commercial systems.

키워드

oxygen consumption ratecellular respiration levelamperometryClark-type sensor
제목
세포 호흡량 측정용 다채널 측정 시스템 개발
제목 (타언어)
Development of a multi channel measurement system for the cellular respiration measurement
저자
남현욱박정일김영미박정호
발행일
2010
저널명
센서학회지
19
1
페이지
36 ~ 42