Endoscopic micro-optical coherence tomography with extended depth of focus using a binary phase spatial filter

  • Kim, Junyoung; 
  • Xing, Jingchao; 
  • Nam, Hyeong Soo; 
  • Song, Joon Woo; 
  • Kim, Jin Won; 
  • 외 1명
Citations

WEB OF SCIENCE

57
Citations

SCOPUS

64

초록

Micro-optical coherence tomography (mu OCT) is an advanced imaging technique that acquires a three-dimensional microstructure of biological samples with a high spatial resolution, up to 1 mu m, by using a broadband light source and a high numerical aperture (NA) lens. As high NA produces a short depth of focus (DOF), extending the DOF is necessary to obtain a reasonable imaging depth. However, due to the complexity of optics and the limited space, it has been challenging to fabricate endoscopic mu OCT, which is essential for clinical translation. Here, we report an endoscopic mu OCT probe with an extended DOF by using a binary phase spatial filter. The imaging results from latex beads demonstrated that the mu OCT probe achieved an axial resolution of 2.49 mu m and a lateral resolution of 2.59 mu m with a DOF extended by a factor of 2. The feasibility of clinical use was demonstrated by ex vivo imaging of the rabbit iliac artery. (C) 2017 Optical Society of America.

키워드

IN-VIVO; SWEPT-SOURCE; ENDOTHELIUM; RANGE
제목
Endoscopic micro-optical coherence tomography with extended depth of focus using a binary phase spatial filter
저자
Kim, Junyoung; Xing, Jingchao; Nam, Hyeong Soo; Song, Joon Woo; Kim, Jin Won; Yoo, Hongki
DOI
10.1364/OL.42.000379
발행일
2017-02-01
유형
Article
저널명
Optics Letters
권
42
호
3
페이지
379 ~ 382