상세 보기
Secondary neutron dose measurement for proton eye treatment using an eye snout with a borated neutron absorber
- Kim, Dong Wook;
- Chung, Weon Kuu;
- Shin, Jungwook;
- Lim, Young Kyung;
- Shin, Dongho;
- ... Yoon, Myongguen;
- 외 4명
WEB OF SCIENCE
12SCOPUS
12초록
Background: We measured and assessed ways to reduce the secondary neutron dose from a system for proton eye treatment. Methods: Proton beams of 60.30 MeV were delivered through an eye-treatment snout in passive scattering mode. Allyl diglycol carbonate (CR-39) etch detectors were used to measure the neutron dose in the external field at 0.00, 1.64, and 6.00 cm depths in a water phantom. Secondary neutron doses were measured and compared between those with and without a high-hydrogen-boron-containing block. In addition, the neutron energy and vertices distribution were obtained by using a Geant4 Monte Carlo simulation. Results: The ratio of the maximum neutron dose equivalent to the proton absorbed dose (H(10)/D) at 2.00 cm from the beam field edge was 8.79 +/- 1.28 mSv/Gy. The ratio of the neutron dose equivalent to the proton absorbed dose with and without a high hydrogen-boron containing block was 0.63 +/- 0.06 to 1.15 +/- 0.13 mSv/Gy at 2.00 cm from the edge of the field at depths of 0.00, 1.64, and 6.00 cm. Conclusions: We found that the out-of-field secondary neutron dose in proton eye treatment with an eye snout is relatively small, and it can be further reduced by installing a borated neutron absorbing material.
키워드
- 제목
- Secondary neutron dose measurement for proton eye treatment using an eye snout with a borated neutron absorber
- 저자
- Kim, Dong Wook; Chung, Weon Kuu; Shin, Jungwook; Lim, Young Kyung; Shin, Dongho; Lee, Se Byeong; Yoon, Myongguen; Park, Sung-Yong; Shin, Dong Oh; Cho, Jung Keun
- 발행일
- 2013-07-17
- 유형
- Article
- 권
- 8