We demonstrated low flow velocity measurement with a miniaturized corona flow sensor operating in drift mobility increment (DMI) mode. The corona flow sensor consisted of a corona flow probe residing in a small diameter flow tube. The corona flow probe had an electrode gap of similar to 700 mu m and was constructed with 50 mu m stainless wire for cathode and nickel plated steel for anode. As evident in both the analytical and experimental corona current versus applied voltage curves, the corona current varied with the ozone concentration in the drift region. This also allowed the corona current to be responsive to air flow. The experimental corona current versus flow velocity trend was shown to be in agreement with the analytical trend. At an applied voltage of 1800V, the corona flow sensor was able to measure flow velocities from 4.7 to 94.3 mm/s with a resolution of similar to 5 mm/s. Three configurations consisting of the corona flow probe in different orientation with respect to the flow direction were investigated. We also showed that the sensitivity and operating range could potentially be tuned by adjusting the applied voltage. Finally, the significance and limitations of the results were also discussed. (C) 2015 Elsevier B.V. All rights reserved.
키워드
- 제목
- Miniaturized corona flow sensor operating in drift mobility increment mode for low flow velocity measurement
- 저자
- Chua, Beelee; Pak, James Jungho
- 발행일
- 2015-04-01
- 유형
- Article
- 권
- 224
- 페이지
- 65 ~ 71
- 언어
- ENG
- 출판사
- ELSEVIER SCIENCE SA
- 발행국가
- 스위스
- 분량
- 7 페이지
- ISSN
- P 0924-4247