Systematic evaluation of recursive approach of EEG-segment-based PCA for removal of helium-pump artefact from MRI

Citations

WEB OF SCIENCE

1
Citations

SCOPUS

2

초록

The cryogenic pump is a crucial component of the magnetic resonance imaging (MRI) system for delivering liquid helium to a magnet for superconductivity, thereby generating a mechanical vibration. Thus, the cryogenic pump for liquid helium (helium pump) contaminates the electroencephalography (EEG) simultaneously acquired with functional MRI. The recursive approach of EEG-segment-based principal component analysis (rsPCA) has recently demonstrated its efficacy in removing this helium pump artefact. In the rsPCA, the recursion depth and EEG-segment size are crucial hyperparameters. rsPCA's performance and computational time across recursion depth (1, 2, 3) and segment sizes (165, 220, 265) are systematically evaluated. It is found that the recursion depth of 2 yielded significant reductions in the computational time compared to the depth of 3 across all segment sizes while maintaining the denoising performance. The binary classification performance (left-hand versus right-hand clenching) was also enhanced in this scenario (especially the use of EEG-segment size of 165 and 220) by using EEG gamma-band activity (30-50 Hz), which is predominantly contaminated by the helium-pump artefact.

키워드

FMRI REVEALS; VALIDATION; REDUCTION; DYNAMICS
제목
Systematic evaluation of recursive approach of EEG-segment-based PCA for removal of helium-pump artefact from MRI
저자
Kim, Hyun-Chul; Lee, Jong-Hwan
DOI
10.1049/ell2.12519
발행일
2022-07
유형
Article
저널명
Electronics Letters
권
58
호
15
페이지
567 ~ 569