초록

In this study, we describe the fabrication of an electrostatic beam position monitor (BPM) for beam diagnostic systems. Calibration of the BPM pickup requires both sensitivity and curve fitting owing to manufacturing tolerances following fabrication. To calibrate the BPM, we developed a BPM calibration system that comprises a wire -based test bench and readout electronics. We considered three difference -over -sum (DOS) data selection methods: DOS diag , DOS axis , and DOS all , to understand their impact on calibration accuracy. Each method selected specific DOS data, and we compared their characteristic curves. Consequently, we analyzed the calibration results by performing linear, 3rd -order, 5th -order, and two-dimensional (2D) 3rd -order polynomial fitting. Each of the three DOS data selection methods was examined to illustrate their advantages and limitations in terms of calibration accuracy and required amount of data. The results obtained demonstrated that choosing the appropriate DOS data -selection approach could be employed for the BPM offline calibrations, achieving minimized calibration errors and optimized response characteristics with polynomial curve fitting.

키워드

Beam-line instrumentation (beam position and profile monitors; beam-intensity monitors; bunch length monitors); Data processing methods; Data reduction methods
제목
Three data-selection approaches for beam position monitor offline calibration
저자
Kim, Geunwoo; Kim, Eun-San; Kwon, Jang Won; Jang, Siwon
DOI
10.1088/1748-0221/19/06/T06006
발행일
2024-06
유형
Article
저널명
Journal of Instrumentation
권
19
호
6