한국인 Cadaver Test에 의한 상완골 골수정 개발에 관한 연구

A Study on Development of Humeral Intramedullary Fixation Nail based on Korean Cadaver Tests
  • 전창수
  • 이재원
  • 고철웅
  • 오종건
  • 우수헌
  • 외 1명

초록

The aim of this study is to develop a humeral Intramedullary fixation nail (HIFN) suitable for Korean people. In this study, CT images were obtained from 72 Korean cadaveric humeral bones and 3D Korean humeral bone models were reconstructed based on the CT images to investigate anatomical characteristics. Major design parameters of HIFN were selected using the morphological measurement information of the Korean humeral bone models. Through finite element analysis and mechanical tests, the developed HIFN prototype was compared with the Polarus HIFN (ACUMEDⓇ, USA), and it was found that the HIFN prototype showed similar and/or superior mechanical performance compared to the Polarus HIFN. Also, clinical validation for the HIFN prototype was carried out to check predictable troubles in surgical operations. Finally,optimal design modification was proposed to prevent the possible axillary nerve injury due to the locking screw system of the HIFN prototype.

키워드

Korean Humerus (한국인 상완골)Intramedullary Fixation Nail (골수정)Finite Element Method (유한요소법)Cadaver Test (사체시험)Axillary Nerve (액와신경)
제목
한국인 Cadaver Test에 의한 상완골 골수정 개발에 관한 연구
제목 (타언어)
A Study on Development of Humeral Intramedullary Fixation Nail based on Korean Cadaver Tests
저자
전창수이재원고철웅오종건우수헌이성재
발행일
2011
저널명
한국정밀공학회지
28
8
페이지
984 ~ 991