Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

In-Situ Nanospectroscopic pH Monitoring by Plasm on Resonance Energy Transfer (PRET)

Authors
Shin, HyunkuBaek, JaeminKwon, HyosungChoi, Yeonho
Issue Date
Nov-2013
Publisher
AMER SCIENTIFIC PUBLISHERS
Keywords
pH Sensor; Gold Nanoparticle; Plasmon Resonance Energy Transfer; Phenol Red; Nanospectroscopy
Citation
JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, v.13, no.11, pp.7287 - 7290
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY
Volume
13
Number
11
Start Page
7287
End Page
7290
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/101670
DOI
10.1166/jnn.2013.8094
ISSN
1533-4880
Abstract
Real-time detection of pH value in a living cell is in central importance for research about cells and diseases. In spite of the great advances in science and technology, pH measurement in a living cell is still limited in spatial resolution, in-situ detection, and intracellular monitoring. Here, we designed a nanoscale pH meter by Plasmon Resonance Energy Transfer (PRET). In order to highly sensitively measure pH with nanoscale spatial resolution, we choose 80 nm spherical gold nanoparticle (GNP) and phenol red which is commonly used in cell media for pH determination. The resonance energy of GNP is transferred to phenol red because the scattering intensity of GNP is overlapped with the second absorption intensity of phenol red at near 560 nm. Meanwhile, the absorption intensity of phenol red molecules is changing with pH value of the solution. For that reason, the intensity of PRET from GNP to phenol red molecules also changes by the acidity of phenol red solution. Then we can detect pH values with nanoscale spatial resolution through the Rayleigh scattering intensity of GNP. As we changed pH value from 6.0 to 9.0, the scattering intensity of GNP is decreased because the absorbance of phenol red at 560 nm wavelength is increased with increasing pH value. The Gaussian peak of a difference in Rayleigh scattering spectra of GNP between pH 6.0 and pH 9.0 indicates exactly the same as UV-vis spectral difference between basic and acidic phenol red solution. We expect that this pH measuring technique has a significant impact on the pH detection of living cells with nanoscale, and it can make possibility to image the cell structure by pH variation.
Files in This Item
There are no files associated with this item.
Appears in
Collections
Graduate School > Department of Bioengineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Choi, Yeon ho photo

Choi, Yeon ho
Department of Bioengineering
Read more

Altmetrics

Total Views & Downloads

BROWSE