Biomechanical assessment of a novel bone lengthening plate system - A cadaveric study

Authors
Jung, Tae GonSuh, Seung WooLee, Sung JaeKim, BongjuHan, Dong-WookYang, Jae Hyuk
Issue Date
2월-2013
Publisher
ELSEVIER SCI LTD
Keywords
Femur; Biomechanical stability; Bone lengthening plate; Bone mineral density
Citation
CLINICAL BIOMECHANICS, v.28, no.2, pp.232 - 238
Indexed
SCIE
SCOPUS
Journal Title
CLINICAL BIOMECHANICS
Volume
28
Number
2
Start Page
232
End Page
238
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/104086
DOI
10.1016/j.clinbiomech.2012.11.011
ISSN
0268-0033
Abstract
Background: Although many types of external fixators have been developed for distraction osteogenesis, all have some drawbacks. We recently developed a novel bone lengthening plate to overcome these problems. The purpose of this study is to conduct biomechanical analyses using cadavers to assess the stability of the bone lengthening plate in relation to distraction length and femoral bone mineral density. Methods: We used human cadaveric femurs (n=18) to assess the effects of distraction length and bone mineral density on the biomechanical stability of the bone lengthening plate. After establishing control (n = 6, 0 mm lengthening) and experimental groups (n = 12, 30 mm lengthening), we measured biomechanical stability (structural stiffness, ultimate load, and displacement) under a compressive load. The experimental group was subdivided into a group with normal bone mineral density (n=6) and a group with osteoporosis (n=6), and the biomechanical stability of these groups was compared. Finding: Structural stiffness differed significantly between the control (417.6 N/mm) and combined experimental groups (185.6 N/mm, p = 0.002). Ultimate load also differed significantly between the control (1327.8 N) and combined experimental (331.4 N, p = 0.002) groups. Bone mineral density was unrelated to structural stiffness (p=0.204), ultimate load (0.876), or displacement (0.344). In all cases, failure of the bone lengthening plate occurred at the longitudinal connectors, such as the connecting columns between the upper and lower plates, and the lengthening shaft of the bone lengthening plate. Interpretation: The biomechanical stability of the bone lengthening plate was affected by the lengthening length but not by bone mineral density. In addition, biomechanical stability during lengthening was most strongly influenced by the longitudinal connectors. (C) 2012 Elsevier Ltd. All rights reserved.
Files in This Item
There are no files associated with this item.
Appears in
Collections
Graduate School > Department of Biomedical Sciences > 1. Journal Articles
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Suh, Seung Woo photo

Suh, Seung Woo
의과학과
Read more

Altmetrics

Total Views & Downloads

BROWSE