Calculation of patient effective dose and scattered dose for dental mobile fluoroscopic equipment: application of the Monte Carlo simulation

Authors
Lee, BoramLee, JungseokKang, SangwonCho, HyelimShin, GwisoonLee, Jeong-WooChoi, Jonghak
Issue Date
1월-2013
Publisher
OXFORD UNIV PRESS
Citation
RADIATION PROTECTION DOSIMETRY, v.153, no.1, pp.80 - 84
Indexed
SCIE
SCOPUS
Journal Title
RADIATION PROTECTION DOSIMETRY
Volume
153
Number
1
Start Page
80
End Page
84
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/104362
DOI
10.1093/rpd/ncs082
ISSN
0144-8420
Abstract
The objective of this study was to evaluate the patient effective dose and scattered dose from recently developed dental mobile equipment in Korea. The MCNPX 2.6 (Los Alamos National Laboratory, USA) was used in a Monte Carlo simulation to calculate both the effective and scattered doses. The MCNPX code was constructed identically as in the general use of equipment and the effective dose and scattered dose were calculated using the KTMAN-2 digital phantom. The effective dose was calculated as 906 Sv. The equivalent doses per organ were calculated via the MCNPX code, and were 32 174 and 19 Sv in the salivary gland and oesophagus, respectively. The scattered dose of 22.532.6 Sv of the tube side at 25 cm from the centre in anterior and posterior planes was measured as 1.43 times higher than the detector side of 10.516.0 Sv.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Health and Environmental Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE