Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Design consideration of a multipinhole collimator with septa for ultra high-resolution silicon drift detector modules

Authors
Min, Byung JunChoi, YongLee, Nam-YongLee, KisungAhn, Young BokJoung, Jinhun
Issue Date
21-7월-2009
Publisher
ELSEVIER
Keywords
Multipinhole collimator; Silicon drift detector; SPECT; Monte Carlo simulation
Citation
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, v.606, no.3, pp.755 - 761
Indexed
SCIE
SCOPUS
Journal Title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
Volume
606
Number
3
Start Page
755
End Page
761
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/119658
DOI
10.1016/j.nima.2009.05.019
ISSN
0168-9002
Abstract
The aim of this study was to design a multipinhole (MP) collimator with lead vertical septa coupled to a high-resolution detector module containing silicon drift detectors (SDDs) with an intrinsic resolution approaching the sub-millimeter level. Monte Carlo simulations were performed to determine pinhole parameters such as pinhole diameter, focal length, and number of pinholes. Effects of parallax error and collimator penetration were investigated for the new MP collimator design. The MP detector module was evaluated using reconstructed images of resolution and mathematical cardiac torso (MCAT) phantoms. In addition, the reduced angular sampling effect was investigated over 180 degrees. The images were reconstructed using dedicated maximum likelihood expectation maximization (MLEM) algorithm. An MP collimator with 81-pinhole was designed with a 2-mm-diameter pinhole and a focal length of 40 mm. Planar sensitivity and resolution obtained using the devised MP collimator were 3.9 cps/mu Ci and 6 mm full-width at half-maximum (FWHM) at a 10 cm distance. The parallax error and penetration ratio were significantly improved using the proposed MP collimation design. The simulation results demonstrated that the proposed MP detector provided enlarged imaging field of view (FOV) and improved the angular sampling effect in resolution and MCAT phantom studies. Moreover, the novel design enables tomography images by simultaneously obtaining eight projections with eight-detector modules located along the 180 degrees orbit surrounding a patient, which allows designing of a stationary cardiac SPECT. In conclusion, the MP collimator with lead vertical septa was designed to have comparable system resolution and sensitivity to those of the low-energy high-resolution (LEHR) collimator per detector. The system sensitivity with an eight-detector configuration would be four times higher than that with a standard dual-detector cardiac SPECT. (C) 2009 Elsevier B.V. All rights reserved.
Files in This Item
There are no files associated with this item.
Appears in
Collections
Graduate School > Department of Bioengineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Ki sung photo

Lee, Ki sung
바이오의공학과
Read more

Altmetrics

Total Views & Downloads

BROWSE