Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis.

Authors
Lim, H.C.Bae, J.H.Hwang, J.H.Kim, S.J.Yoon, J.Y.
Issue Date
2009
Citation
Clinics in orthopedic surgery, v.1, no.2, pp.118 - 121
Indexed
SCOPUS
Journal Title
Clinics in orthopedic surgery
Volume
1
Number
2
Start Page
118
End Page
121
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/121850
DOI
10.4055/cios.2009.1.2.118
ISSN
2005-291x
Abstract
We report the case of a polyethylene tibial post fracture in a 72-year-old woman 14 months after a Scorpio posterior-stabilized (PS) total knee arthroplasty. The polyethylene wear was found around the fracture site of the post, especially over the anterior aspect of the post base. The failure mechanism of the post fracture in the present case was anterior impingement with excessive wear over the base of the anterior aspect of the tibial post, which became a stress-riser of post and cam articulation. This is the first report of a polyethylene tibial post fracture of a Scorpio PS prosthesis.