Performance of ivisen IA-1400, a new point-of-care device with an internal centrifuge system, for the measurement of cardiac troponin I levels

Authors
Kim, Ha NuiYoon, Soo-Young
Issue Date
5월-2021
Publisher
WILEY
Keywords
cardiac troponin; cardiac troponin I; i& #8208; SENS; ivisen IA& #8208; 1400; point& #8208; of& #8208; care
Citation
JOURNAL OF CLINICAL LABORATORY ANALYSIS, v.35, no.5
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF CLINICAL LABORATORY ANALYSIS
Volume
35
Number
5
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/137420
DOI
10.1002/jcla.23747
ISSN
0887-8013
Abstract
Background We present the analytical performance of the ivisen IA-1400, a new point-of-care device that features a characteristic built-in centrifuge system, to measure blood cardiac troponin I (cTnI) levels. Methods Whole blood and plasma samples obtained from patients who visited Korea University Guro Hospital were used to analyze measurement range, cross-reactivity, interference, and sensitivity and specificity. We performed a correlation analysis of the ivisen IA-1400 versus the Access AccuTnI+3 immunoassay using the UniCel (TM) DxI 800 platform and the PATHFAST (TM) hs-cTnI assay. Results Within-run precisions were classified as low, 9.8%; middle, 10.2%; and high, 8.5%. The limit of blank was 3.1 ng/L for plasma samples and 4.3 ng/L for whole blood samples. The limit of detection was 8.4 ng/L for plasma samples and 10.0 ng/L for whole blood samples, respectively. The limit of quantitation at a coefficient of variation of 20% and 10% was 19.5 ng/L and 45.5 ng/L for plasma samples, respectively. The comparative evaluation between the two other assays and ivisen IA-1400 showed excellent correlation, with Spearman's correlation coefficients (R) of 0.992 and 0.985. The sensitivity and specificity of ivisen IA-1400 using the optimum cut-off value of 235 ug/L were 94.6% and 98.2%, respectively. Conclusion The ivisen IA-1400 showed acceptable and promising performance in cTnI measurements using whole blood and plasma samples, with limited information in the clinical performance. The flexibility for sample selection using the internal centrifugation system is the main advantage of this point-of-care device.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Soo Young photo

Yoon, Soo Young
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE