Performance of ivisen IA-1400, a new point-of-care device with an internal centrifuge system, for the measurement of cardiac troponin I levels
- Authors
- Kim, Ha Nui; Yoon, Soo-Young
- Issue Date
- 5월-2021
- Publisher
- WILEY
- Keywords
- cardiac troponin; cardiac troponin I; i& #8208; SENS; ivisen IA& #8208; 1400; point& #8208; of& #8208; care
- Citation
- JOURNAL OF CLINICAL LABORATORY ANALYSIS, v.35, no.5
- Indexed
- SCIE
SCOPUS
- Journal Title
- JOURNAL OF CLINICAL LABORATORY ANALYSIS
- Volume
- 35
- Number
- 5
- URI
- https://scholar.korea.ac.kr/handle/2021.sw.korea/137420
- DOI
- 10.1002/jcla.23747
- ISSN
- 0887-8013
- Abstract
- Background We present the analytical performance of the ivisen IA-1400, a new point-of-care device that features a characteristic built-in centrifuge system, to measure blood cardiac troponin I (cTnI) levels. Methods Whole blood and plasma samples obtained from patients who visited Korea University Guro Hospital were used to analyze measurement range, cross-reactivity, interference, and sensitivity and specificity. We performed a correlation analysis of the ivisen IA-1400 versus the Access AccuTnI+3 immunoassay using the UniCel (TM) DxI 800 platform and the PATHFAST (TM) hs-cTnI assay. Results Within-run precisions were classified as low, 9.8%; middle, 10.2%; and high, 8.5%. The limit of blank was 3.1 ng/L for plasma samples and 4.3 ng/L for whole blood samples. The limit of detection was 8.4 ng/L for plasma samples and 10.0 ng/L for whole blood samples, respectively. The limit of quantitation at a coefficient of variation of 20% and 10% was 19.5 ng/L and 45.5 ng/L for plasma samples, respectively. The comparative evaluation between the two other assays and ivisen IA-1400 showed excellent correlation, with Spearman's correlation coefficients (R) of 0.992 and 0.985. The sensitivity and specificity of ivisen IA-1400 using the optimum cut-off value of 235 ug/L were 94.6% and 98.2%, respectively. Conclusion The ivisen IA-1400 showed acceptable and promising performance in cTnI measurements using whole blood and plasma samples, with limited information in the clinical performance. The flexibility for sample selection using the internal centrifugation system is the main advantage of this point-of-care device.
- Files in This Item
- There are no files associated with this item.
- Appears in
Collections - College of Medicine > Department of Medical Science > 1. Journal Articles
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.