Simultaneous point-of-care testing of blood lipid profile and glucose: Performance evaluation of the GCare Lipid Analyzer

Authors
Kim, Ha NuiYoon, Soo-Young
Issue Date
12월-2021
Publisher
WILEY
Keywords
diabetes mellitus; dyslipidemia; glucose; lipid; POC; point-of-care testing
Citation
JOURNAL OF CLINICAL LABORATORY ANALYSIS, v.35, no.12
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF CLINICAL LABORATORY ANALYSIS
Volume
35
Number
12
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/140520
DOI
10.1002/jcla.24055
ISSN
0887-8013
Abstract
Background Point-of-care (POC) testing provides quick results and includes tests for blood glucose and lipid profiles. We evaluated the newly developed POC device, the GCare Lipid Analyzer, which is used to measure glucose, total cholesterol (TC), triglyceride (TG), and high-density lipoprotein cholesterol (HDL-C) levels. Methods Venous and capillary blood samples were collected from patients who visited Korea University Guro Hospital. The results obtained using the GCare Lipid Analyzer were compared with those obtained using the TBA 2000FR chemistry analyzer and YSI 2300 STAT Plus analyzer. The glucose system evaluation process was based on the International Organization for Standardization 15197:2013 guidelines. Results The correlation coefficients (R) for TC, TG, and HDL-C were 0.965, 0.969, and 0.943 in capillary blood and 0.969, 0.990, and 0.956 in venous blood, respectively. The total errors for TC, TG, and HDL-C of the lipid profile using venous blood were all acceptable at 6.6%, 9.3%, and 11.6%, respectively. For glucose concentrations <100 mg/dl, 96.1% of the measured glucose levels were within +/- 15 mg/dl in venous samples and 100% were within +/- 15 mg/dl in capillary samples. For glucose concentrations >= 100 mg/dl, 100% and 99.5% of the measured glucose levels were within 15% for venous and capillary blood, respectively. Conclusion The performance of the GCare Lipid Analyzer is acceptable for both blood glucose and lipid profile testing, indicating that it is reliable for use in patients with diabetic dyslipidemia.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Soo Young photo

Yoon, Soo Young
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE