To develop a second set of diagnostic reference levels (DRLs) and achievable doses (ADs) for 13 adult computed tomography (CT) protocols and a paediatric head CT protocol in Korea. A survey of 13,625 CT examinations was performed based on 13 adult CT protocols and a paediatric non-contrast brain CT protocol using 369 CT systems, with patients grouped according to age. Most CT protocols in this survey had DRLs similar to those reported in other countries. However, chest and abdomen-pelvic CT had lower DRLs than those reported in the first Korean national survey and those from other countries. Paediatric non-contrast brain CT in each age group, with the exception of the 11-15-year age group, had lower DRLs than those reported in other countries. The DRLs presented here are similar to (or lower than for some protocols) those reported in the first Korean national survey and those from other countries.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Health and Environmental Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE