Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

복합레진으로 제작한 인레이 보철물 구조에 따른 교합면 부위의 2차원 변연 적합도 및 내면 부위의 3차원 정확성 분석Analysis of the 2-dimensional marginal fit of the occlusal surface and the 3-dimensional accuracy of the inner surface of the occlusal surface according to the inlay prosthesis structure made of composite resin

Other Titles
Analysis of the 2-dimensional marginal fit of the occlusal surface and the 3-dimensional accuracy of the inner surface of the occlusal surface according to the inlay prosthesis structure made of composite resin
Authors
김웅철김동연이태희박동인박진영정일도이하나김지환
Issue Date
2019
Publisher
대한치과기공학회
Keywords
Marginal fit; Accuracy; 3-Dimension; Inlay; Composite resin
Citation
대한치과기공학회지, v.41, no.1, pp.21 - 27
Indexed
KCI
Journal Title
대한치과기공학회지
Volume
41
Number
1
Start Page
21
End Page
27
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/69629
ISSN
1229-3954
Abstract
Purpose: To evaluate 2D and 3D of occulsal, mesial-occlusal and mesial-occlusal-distal cavity of composite resin inlay. Methods: Abutment tooth 16, 36 of FDI system was selected for the study. Inlay prostheses classified as occlusal cavity (OC group), mesial-occlusal (MOC) and mesial-occlusal-distal cavity (MODC) were prepared using composite resin. Composite resin was injected with composite resin in prepared tooth cavity and then photopolymerized with UV light. Additional thermal polymerization was performed. Marginal gap of composite resin inlays were measured by digital microscope(x160) with silicone replica technique. The data was analyzed from statistical software for Kruskal-Wallis test (α=0.05). 3-dimensional analysis was analyzed through superimposition method. Results: The smallest 2D marginal fit measure of the three groups was 47.0±21.6 μm in the MOC group. The largest 2D marginal was 69.1 ± 33.8 μm in the MODC group. In the trueness of the three groups, the most accurate figure was 14.4±2.3 μm for the MODC group. In Precision, the most accurate figure was 14.5±4.3 μm for the MODC group. Conclusion : In this study, 2D marginal fit of OC, MOC, and MODC cavities fabricated with composite resin was applicable to all clinical applications. In the 3D inner surface accuracy evaluation, the MODC group showed the accuracy results.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > Division of Health Policy and Management > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Ji Hwan photo

Kim, Ji Hwan
보건과학대학 (보건정책관리학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE