Effects of Varying Matching Between Collimator Hole and Scintillator Pixel on Gamma Camera Image Resolution

Authors
Kim, MinhoBae, Jae KeonKim, Kyeong MinLee, Wonho
Issue Date
2월-2018
Publisher
KOREAN PHYSICAL SOC
Keywords
Gamma camera; Pixelated scintillator; Parallel hole collimator; Monte Carlo simulation; Spatial resolution
Citation
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, v.72, no.3, pp.455 - 461
Indexed
SCIE
SCOPUS
KCI
Journal Title
JOURNAL OF THE KOREAN PHYSICAL SOCIETY
Volume
72
Number
3
Start Page
455
End Page
461
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/77416
DOI
10.3938/jkps.72.455
ISSN
0374-4884
Abstract
A collimator is a key component of gamma cameras, which affects imaging performance significantly. Recently, pixelated scintillator arrays are widely applied in the development of highresolution gamma cameras, instead of monolithic scintillator blocks. Therefore, combining the collimator and the pixelated scintillator has been a crucial factor in achieving good imaging performance. In this study, we investigated the effects of the shape and position of a collimator hole on the image resolution of a gamma camera with a pixelated scintillator. We designed a pixel-matched collimator and a low-energy high-resolution (LEHR) collimator with hexagonal shaped holes that have the same collimator geometrical efficiency. Simulation experiments were performed with a Monte Carlo simulation package (GATE v7.0). For both types of collimator, camera sensitivity was measured for validating the collimator efficiency, and planar images were obtained to measure spatial resolution, with changing the line profile angle based on the septa direction. Camera sensitivities for both collimators were equivalent (difference = 1.76 +/- 0.51%). In contrast to the case of the pixel-matched collimator, in the case of the LEHR collimator, as the angle of the line profile varied, the spatial resolution showed a difference at each profile angle. In the case when two collimator holes symmetrically matched with one scintillator pixel, the spatial resolution showed a 64.87% difference in its maximum and a position distortion of 26.06% from the original source position, according to the profile angle, compared to the other unmatched cases. These results showed that using a pixel-matched collimator in gamma camera systems employing a pixelated scintillator is essential for the exact delineation of small regions using high resolution gamma camera imaging.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Health and Environmental Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Lee, Won ho photo

Lee, Won ho
보건과학대학 (보건환경융합과학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE