Biomechanical function of a balloon nucleus pulposus replacement system: A human cadaveric spine study

Authors
Lee, TaekyeongLim, Tae-HongLee, Sang-HeonKim, Joo-HanHong, Junghwa
Issue Date
1월-2018
Publisher
WILEY
Keywords
disc nucleus replacement; intervertebral disc; biomechanical functions; balloon nucleus replacement system
Citation
JOURNAL OF ORTHOPAEDIC RESEARCH, v.36, no.1, pp.167 - 173
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF ORTHOPAEDIC RESEARCH
Volume
36
Number
1
Start Page
167
End Page
173
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/78519
DOI
10.1002/jor.23607
ISSN
0736-0266
Abstract
With recent advances in motion-sparing techniques in spine surgery, disc nucleus replacement (DNR) has been introduced as a viable method to restore the biomechanical functions of the spine. Several methods of DNR have been proposed in the literature. However, the risk of device migration or extrusion is a major issue that should be addressed for a successful DNR. DNR using a balloon nucleus (BN) filled with pressurized fluid may be capable of reducing such risks while preserving the advantages of DNR. The objective of this study was to investigate the biomechanical functionalities of the human cadaveric lumbar motion segments with a custom made BN filled with saline at internal fluid pressure of 0.3 or 0.6MPa in terms of axial and rotational flexibilities of the L4-L5 motion segment. Axial flexibility was quantified by the axial displacement resulting from an axial compressive force of 400N while the rotational flexibility by the range of motions determined as the rotational angles in response to a pure moment of 6.0Nm in flexion, extension, and right- and left-lateral bending directions. These tests were performed successively on the motion segment in the following conditions: intact, post nucleotomy, implanting BN with 0.3MPa, and BN with 0.6MPa. The nucleotomy was found to significantly increase both the axial and rotational flexibilities while the implantation of the BN reduced the axial and rotational flexibilities to those of the intact segment. The axial and rotational flexibilities of the segment with the BN with 0.3MPa were greater than those of the segment with the BN with 0.6MPa. (c) 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 36:167-173, 2018.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles
Graduate School > Department of Control and Instrumentation Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher HONG, JUNG HWA photo

HONG, JUNG HWA
제어계측공학과
Read more

Altmetrics

Total Views & Downloads

BROWSE