Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Fully implantable, battery-free wireless optoelectronic devices for spinal optogenetics

Authors
Samineni, Vijay K.Yoon, JangyeolCrawford, Kaitlyn E.Jeong, Yu RaMcKenzie, Kajanna C.Shin, GunchulXie, ZhaoqianSundaram, Saranya S.Li, YuhangYang, Min YoungKim, JeonghyunWu, DiXue, YeguangFeng, XueHuang, YonggangMickle, Aaron D.Banks, AnthonyHa, Jeong SookGolden, Judith P.Rogers, John A.Gereau, Robert W.
Issue Date
11월-2017
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Keywords
Optogenetics; Wireless; Spinal; Peripheral; NFC; Optoelectronic; Dorsal horn; Freely moving; ChR2; NpHr3.0; Channelrhodopsin; Halorhodopsin; Archaerhodopsin; ArchT; Light; Biocompatible; Optogenetic; PAIN; Itch
Citation
PAIN, v.158, no.11, pp.2108 - 2116
Indexed
SCIE
SCOPUS
Journal Title
PAIN
Volume
158
Number
11
Start Page
2108
End Page
2116
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/81710
DOI
10.1097/j.pain.0000000000000968
ISSN
0304-3959
Abstract
The advent of optogenetic tools has allowed unprecedented insights into the organization of neuronal networks. Although recently developed technologies have enabled implementation of optogenetics for studies of brain function in freely moving, untethered animals, wireless powering and device durability pose challenges in studies of spinal cord circuits where dynamic, multidimensional motions against hard and soft surrounding tissues can lead to device degradation. We demonstrate here a fully implantable optoelectronic device powered by near-field wireless communication technology, with a thin and flexible open architecture that provides excellent mechanical durability, robust sealing against biofluid penetration and fidelity in wireless activation, thereby allowing for long-term optical stimulation of the spinal cord without constraint on the natural behaviors of the animals. The system consists of a double-layer, rectangular-shaped magnetic coil antenna connected to a microscale inorganic light-emitting diode (m-ILED) on a thin, flexible probe that can be implanted just above the dura of the mouse spinal cord for effective stimulation of light-sensitive proteins expressed in neurons in the dorsal horn. Wireless optogenetic activation of TRPV1-ChR2 afferents with spinal m-ILEDs causes nocifensive behaviors and robust real-time place aversion with sustained operation in animals over periods of several weeks to months. The relatively low-cost electronics required for control of the systems, together with the biocompatibility and robust operation of these devices will allow broad application of optogenetics in future studies of spinal circuits, as well as various peripheral targets, in awake, freely moving and untethered animals, where existing approaches have limited utility.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Engineering > Department of Chemical and Biological Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Ha, Jeong Sook photo

Ha, Jeong Sook
공과대학 (화공생명공학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE