Implant Drill Characteristics: Thermal and Mechanical Effects of Two-, Three-, and Four-Fluted Drills

Authors
Oh, Hyun JunKim, Beom IkKim, Hae-YoungYeo, In-SungWikesjo, Ulf M. E.Koo, Ki-Tae
Issue Date
5월-2017
Publisher
QUINTESSENCE PUBLISHING CO INC
Keywords
cutting efficiency; drill geometry; frictional heat; implant drill design
Citation
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, v.32, no.3, pp.483 - 488
Indexed
SCIE
SCOPUS
Journal Title
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS
Volume
32
Number
3
Start Page
483
End Page
488
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/83649
DOI
10.11607/jomi.4819
ISSN
0882-2786
Abstract
Purpose: Avoiding excessive trauma-thermal or otherwise-during dental implant site preparation is considered critical to implant success; overheating is considered to be a major cause of bone necrosis. Studies evaluating thermal and mechanical effects of implant drill design are limited, and effects of flute design have not been accounted for. The purpose of this study was to compare heat generation and cutting efficiency associated with two-, three-, and four-fluted implant drills to investigate the optimal number of flutes. Materials and Methods: Two-, three-, and four-fluted dental implant drills with identical point, relief, and rake angles and otherwise standard dimensions were evaluated. Real-time temperature changes while drilling artificial bone were recorded using an infrared thermal imager. Cutting efficiency was assessed as the drilling time to a 15-mm depth under constant load using a specially designed recording system. Each drill variation was examined 20 times. A one-way analysis of variance was used for statistical analysis. Results: Mean temperature increases amounted to 8.3 degrees C, 10.8 degrees C, and 15.1 degrees C for two-, three-, and four-fluted drills, respectively; temperatures significantly increased (P < .001) with an increased number of flutes. Mean drilling time serving as a measure of cutting efficiency amounted to 2.6, 2.5, and 2.5 seconds for the two-, three-, and four-fluted drills, respectively. A trend of cutting efficiency increasing or decreasing according to the number of flutes was not observed. Differences in cutting efficiency among the three drill variations were statistically significant (P = .015). The cutting efficiency of the three-fluted drill was superior to that of the two-fluted drill (P = .016). Conclusion: Within the limitations of the study, a two-fluted drill would be preferred for osteotomy preparation due to its level of heat generation, whereas a three-fluted drill showed favorable cutting efficiency.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > Division of Health Policy and Management > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Hae Young photo

Kim, Hae Young
보건과학대학 (보건정책관리학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE