Detailed Information

Cited 0 time in webofscience Cited 0 time in scopus
Metadata Downloads

Optimal stimulation during monitored thyroid surgery: EMG response characteristics in a porcine model

Authors
Wu, Che-WeiLiu, XiaoliBarczynski, MarcinKim, Hoon YubDionigi, GianlorenzoSun, HuiChiang, Feng-YuKamani, DiptiRandolph, Gregory W.
Issue Date
4월-2017
Publisher
WILEY
Keywords
Intraoperative neural monitoring (IONM); thyroid and parathyroid surgery; external branch of superior laryngeal nerve (EBSLN); recurrent laryngeal nerve (RLN); vagus nerve (VN); monopolar probes; bipolar probes; stimulation dissectors
Citation
LARYNGOSCOPE, v.127, no.4, pp.998 - 1005
Indexed
SCIE
SCOPUS
Journal Title
LARYNGOSCOPE
Volume
127
Number
4
Start Page
998
End Page
1005
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/83803
DOI
10.1002/lary.26141
ISSN
0023-852X
Abstract
ObjectivesTo compare electromyography (EMG) characteristics of the external branch of superior laryngeal nerve (EBSLN), recurrent laryngeal nerve (RLN), and vagus nerve (VN) evoked with different stimulation probes/dissectors during monitored thyroid surgery. Study DesignExperimental porcine model. MethodsIn five piglets (10 EBSLNs/RLNs/VNs), laryngeal EMG was recorded by endotracheal tube surface electrodes with stimulation using five monopolar probes (group I), three bipolar probes (group II), and two stimulation dissectors (group III). The detectable EMG response (DER) was defined as > 100 V and was obtained with these different probes/dissectors. Electromyography parameters, stimulus-response curve, and distance-sensitivity results were compared. ResultsAll stimulation probes/dissectors evoked typical EMG waveforms from the EBSLN/RLN/VN with 1 mA current. A stimulus-response curve with increasing EMG amplitude with increase in stimulating current was noted, with the maximum EMG elicited by group I/III probes/dissectors at < 1 mA and at a higher current for group II probes. All groups recorded lower evoked EMG amplitudes when the nerve was stimulated with overlying fascia or when probe/dissector to nerve distance was greater. The mean amplitude decreased by 11%/33%/13% in group I/II/III probes/dissectors when stimulating nerves covered by fascia versus nerves dissected free of overlying fascia. The rate of obtaining DER at 1- or 2-mm distance was significantly higher in group I than in group II/III probes/dissectors (P < 0.001). Latency did not change with any of the stimulation probes/dissectors or trials. ConclusionMonopolar, bipolar probes, and newer stimulation dissectors all provided valid evoked VN/RLN/EBSLN waveforms. They have different functional sensitivity profiles and vary when stimulating with fascia and at a distance from the nerve. Selection of a stimulation probe/dissector for nerve monitoring can be based on the stimulation characteristics, the intended nerve monitoring application, and the surgeon's preference. Level of EvidenceN/A. Laryngoscope, 127:998-1005, 2017
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Hoon Yub photo

Kim, Hoon Yub
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE