Effect of Implant Diameter on Fatigue Strength

Authors
Song, So-YeonLee, Jeong-YolShin, Sang-Wan
Issue Date
2월-2017
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Keywords
narrow-diameter implants; mini-implants; smaller-diameter implants; dynamic fatigue test; cyclic load test
Citation
IMPLANT DENTISTRY, v.26, no.1, pp.59 - 65
Indexed
SCIE
SCOPUS
Journal Title
IMPLANT DENTISTRY
Volume
26
Number
1
Start Page
59
End Page
65
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/84745
DOI
10.1097/ID.0000000000000502
ISSN
1056-6163
Abstract
Purpose: To investigate the effect of implant diameter on fatigue strength using static and cyclic load test. Materials and Methods: Four different implant systems-SuperLine (Phi 4.0), NRLine (Phi 3.1), SlimLine (Phi 2.8, Phi 2.3), and (Dentium) dwere grouped by implant diameter. A static load test was conducted with 5 specimens for each group using a universal testing machine to measure the ultimate failure load (UFL). With 80% of the UFL in the weakest group, the starting load for a cyclic load test was determined and the test was performed with 8 specimens for each group. All tests were conducted according to ISO14801 (2007) until implant failure occurred. After dynamically loaded, each specimen was sectioned and stereo-microscopically examined. The failure modes of each implant system were classified. Static and cyclic load test data were respectively analyzed after the test of normality, with the level of significance at P = 0.05. Results: In the static load test, the higher maximum load of the standard-diameter implant was significant compared with the recorded narrow or mini-implants (P, 0.05). The yield strengths of the Phi 2.8 and Phi 3.1 implants were significantly greater than that of the Phi 2.3 implant (P < 0.05). In a cyclic load test, the mean number of cycles until implant failure occurred was recorded for each specimen. The value for the Phi 4.0 implant was significantly greater (P < 0.001). Conclusion: Implant diameter has an effect on the ability to withstand both static and cyclic loads within Dentium implant systems, The UFLs and fatigue cycles decreased as the implants diameter became smaller.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE