Electrophysiologic monitoring correlates of recurrent laryngeal nerve heat thermal injury in a porcine model

Authors
Lin, Yi-ChuDionigi, GianlorenzoRandolph, Gregory W.Lu, I-ChengChang, Pi-YingTsai, Shan-YinKim, Hoon YubLee, Hye YoonTufano, Ralph P.Sun, HuiLiu, XiaoliChiang, Feng-YuWu, Che-Wei
Issue Date
8월-2015
Publisher
WILEY
Keywords
Recurrent laryngeal nerve; continuous intraoperative neuromonitoring; thyroid surgery; nerve thermal injury; energy-based devices
Citation
LARYNGOSCOPE, v.125, no.8, pp.E283 - E290
Indexed
SCIE
SCOPUS
Journal Title
LARYNGOSCOPE
Volume
125
Number
8
Start Page
E283
End Page
E290
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/92822
DOI
10.1002/lary.25362
ISSN
0023-852X
Abstract
Objectives/HypothesisThermal injury to the recurrent laryngeal nerve (RLN) may not be visually apparent and may go unrecognized intraoperatively. This study aimed to investigate the heat thermal tolerance of RLN and evaluate the electrophysiologic correlates of electromyographic (EMG) signal change during an acute RLN heat damage. Study DesignProspective porcine model with continuous intraoperative neuromonitoring (CIONM). MethodsTen pigs (20 RLNs) undergoing CIONM had their EMG tracings recorded and correlated with heated normal saline (NS) irrigation of varying temperature and duration. ResultsIn the initial pilot study, the EMG was without change during incremental heated NS irrigation (40/45/50/55 degrees C for 60 seconds), but adverse EMG combined events (CE) (amplitude decrease with a concordant latency increase) occurred and degraded to loss of signal (LOS) (by 17.51.3 seconds) when the temperature was elevated to 60 degrees C (n=4). Another 16 RLNs were evaluated to further compare the EMG pattern after various degrees of thermal stress (60/70 degrees C for 30/20 seconds). Electromyographic recordings showed CEs and LOS in all RLNs, and only six of eight RLNs with 60 degrees C exposure showed slight EMG amplitude recovery (16%-35%) after 20 minutes. None of the injured nerve segments were visually apparent, but all were detectable by IONM. ConclusionSixty degrees Celsius is a critical temperature to cause RLN thermal injury. Continuous intraoperative neuromonitoring can be used as a tool for the early detection of acute thermal stress and may guide use of energy-based devices during thyroid procedures. Level of EvidenceN/A. Laryngoscope, 125:E283-E290, 2015
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Hoon Yub photo

Kim, Hoon Yub
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE