Comparison of Intraocular Pressure Measurements between Icare Pro Rebound Tonometer and Tono-Pen XL Tonometer in Supine and Lateral Decubitus Body Positions

Authors
Lee, Tae-EunYoo, ChungkwonHwang, Jin-YoungLin, ShanKim, Yong Yeon
Issue Date
2015
Publisher
TAYLOR & FRANCIS INC
Keywords
Body position; Icare tonometer; intraocular pressure; lateral decubitus; rebound tonometer; supine; tono-pen tonometer
Citation
CURRENT EYE RESEARCH, v.40, no.9, pp.923 - 929
Indexed
SCIE
SCOPUS
Journal Title
CURRENT EYE RESEARCH
Volume
40
Number
9
Start Page
923
End Page
929
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/96240
DOI
10.3109/02713683.2014.964416
ISSN
0271-3683
Abstract
Purpose: To compare intraocular pressure (IOP) measurements obtained using the Icare Pro rebound tonometer and Tono-Pen XL tonometer in supine and lateral decubitus body positions. Methods: One-hundred eyes of 50 subjects (normal volunteers or glaucoma suspects) were enrolled in this prospective observational study. IOP was measured in both eyes using the Icare Pro and Tono-Pen XL in the sitting position and the recumbent positions including supine, right lateral decubitus and left lateral decubitus. IOP was measured five minutes after assuming each of the recumbent postures in a randomized sequence. The eye on the lower side in the lateral decubitus position was termed as the dependent eye. Agreement of IOP readings between the Icare Pro and Tono-Pen was assessed in all recumbent positions. Differences of IOP readings (Delta IOP) between the two tonometers and their correlations with ocular parameters were also assessed in all positions. Results: The IOP readings obtained using Icare Pro and Tono-Pen showed good correlations in supine and lateral decubitus positions (all r>0.7, p<0.005), although Icare Pro readings were higher than Tono-Pen readings (all p<0.001) in all positions. The DIOP showed a weakly positive correlation with central corneal thickness in both eyes, whereas such a positive correlation was found only in the dependent eye in the lateral decubitus positions (r=0.307-0.531, all p<0.005). Both the spherical equivalents and axial lengths were not correlated with Delta IOP in all positions. Conclusion: IOP readings obtained with Tono-Pen and Icare Pro tonometers showed good agreement in supine as well as in lateral decubitus positions, although Icare readings were higher than Tono-Pen readings in all positions. Such differences in IOP readings between the different tonometers need to be considered when measuring IOPs in various body positions.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoo, Chung kwon photo

Yoo, Chung kwon
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE