Comparison of chest radiography, chest digital tomosynthesis and low dose MDCT to detect small ground-glass opacity nodules: an anthropomorphic chest phantom study

Authors
Doo, Kyung WonKang, Eun-YoungYong, Hwan SeokHam, Soo-YounLee, Ki YeolChoo, Ji Yung
Issue Date
12월-2014
Publisher
SPRINGER
Keywords
Pulmonary nodule; Digital tomosynthesis; Computed tomography; Thoracic imaging; Radiography
Citation
EUROPEAN RADIOLOGY, v.24, no.12, pp.3269 - 3276
Indexed
SCIE
SCOPUS
Journal Title
EUROPEAN RADIOLOGY
Volume
24
Number
12
Start Page
3269
End Page
3276
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/96605
DOI
10.1007/s00330-014-3376-6
ISSN
0938-7994
Abstract
The purpose of this study was to evaluate the diagnostic performance of chest radiography (CXR), chest digital tomosynthesis (DT) and low dose multidetector computed tomography (LDCT) for the detection of small pulmonary ground-glass opacity (GGO) nodules, using an anthropomorphic chest phantom. Artificial pulmonary nodules were placed in a phantom and a total of 40 samples of different nodule settings underwent CXR, DT and LDCT. The images were randomly read by three experienced chest radiologists. Free-response receiver-operating characteristics (FROC) were used. The figures of merit for the FROC curves averaged for the three observers were 0.41, 0.37 and 0.76 for CXR, DT and LDCT, respectively. FROC analyses revealed significantly better performance of LDCT over CXR or DT for the detection of GGO nodules (P < 0.05). The difference in detectability between CXR and DT was not statistically significant (P = 0.73). The diagnostic performance of DT for the detection of pulmonary small GGO nodules was not significantly different from that of CXR, but LDCT performed significantly better than both CXR and DT. DT is not a suitable alternative to CT for small GGO nodule detection, and LDCT remains the method of choice for this purpose. aEuro cent For GGO nodule detection, DT was not significantly different from CXR. aEuro cent DT is not a suitable alternative to CT for GGO nodule detection. aEuro cent LDCT is the method of choice for GGO nodule detection.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yong, Hwan Seok photo

Yong, Hwan Seok
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE