EVALUATION OF THE ACCURACY OF DOSE DELIVERY FOR IMRT BASED ON TRANSIT DOSIMETRY

Authors
Baek, Tae SeongChung, Eun JiKoh, Eui KwanSeo, JungjuYoon, Myonggeun
Issue Date
9월-2014
Publisher
LIPPINCOTT WILLIAMS & WILKINS
Keywords
exposure, radiation; physics, medical; quality assurance; radiation, medical
Citation
HEALTH PHYSICS, v.107, no.3, pp.200 - 205
Indexed
SCIE
SCOPUS
Journal Title
HEALTH PHYSICS
Volume
107
Number
3
Start Page
200
End Page
205
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/97575
DOI
10.1097/HP.0000000000000098
ISSN
0017-9078
Abstract
The authors have developed and evaluated a new method that uses transit dose in the treatment room in verification of dose delivery to patients. Five intensity modulated radiotherapy (IMRT) plans were selected from actual cancer patients, and transit dose was measured using MapCHECK 2 and an anthropomorphic phantom. The criteria used as a verification tool for the dose delivery to the patient were gamma-index-based dose comparison between the computed dose and measured dose. When the doses were delivered to an anthropomorphic phantom normally, the average passing rate was 95.2% based on a gamma index analysis. This feasibility study suggested that transit dose-based quality assurance can provide information about the accuracy of an inhomogeneity correction algorithm and patient positioning during treatment, allowing its use as a verification tool for actual dose delivery to patients in the treatment room.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Health Sciences > School of Biomedical Engineering > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Yoon, Myong geun photo

Yoon, Myong geun
보건과학대학 (바이오의공학부)
Read more

Altmetrics

Total Views & Downloads

BROWSE