A simple method to shorten the unfolding time of prehydrated hydrophobic intraocular lens

Authors
Eom, YoungsubLee, Joon SikRhim, Jay WonKang, Su-YeonSong, Jong SukKim, Hyo Myung
Issue Date
8월-2014
Publisher
CANADIAN OPHTHAL SOC
Citation
CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, v.49, no.4, pp.382 - 387
Indexed
SCIE
SCOPUS
Journal Title
CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE
Volume
49
Number
4
Start Page
382
End Page
387
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/97803
DOI
10.1016/j.jcjo.2014.06.002
ISSN
0008-4182
Abstract
Objective: To evaluate the effect of warm ophthalmic viscosurgical devices (OVDs) on the unfolding time of prehydrated hydrophobic acrylic intraocular lenses (IOLs). Design: Experimental study and human trial. Participants: Three foldable hydrophobic acrylic IOLs (en Vista MX60, AcrySof SN60AT, and Tecnis 1 ZCB00). Methods: The unfolding times of 3 kinds of IOLs were measured according to temperature from 26 degrees C to 32 degrees C in a transparent container filled with a mixture of OVDs and balanced salt solution. The unfolding time of each IOL was measured 4 times for each temperature. Unfolding time was defined as the time required for the folded IOL to recover >= 90% of its overall optic diameter before folding. In human trials, the unfolding time of the MX60 in a capsular bag filled with 30 degrees C OVDs was compared with that filled with room temperature OVDs for 4 cases in each group. Results: The unfolding time of the MX60 (215 +/- 25 seconds) was significantly longer than that of the SN60AT (28 +/- 7 seconds) and the ZCBOO (29 +/- 7 seconds) at 26 degrees C (p = 0.013). However, there were no differences in the unfolding time of 3 IOLs at 32 degrees C. In human trials, the unfolding time of the MX60 was shorter in a capsular bag filled with 30 degrees C OVDs (32 13 seconds) than if filled with OVDs kept at room temperature (127 27 seconds; p = 0.029). Conclusions: When fast and complete unfolding characteristics are needed, filling the anterior chamber and capsular bag with OVDs warmed to 30 degrees C before IOL implantation is recommended.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Song, Jong Suk photo

Song, Jong Suk
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE