Oxford phase III meniscal bearing fracture: Case report

Authors
Lim, Hong-ChulShon, Won-YongKim, Seung-JuBae, Ji-Hoon
Issue Date
1월-2014
Publisher
ELSEVIER SCIENCE BV
Keywords
Oxford unicompartmental; Knee; Meniscal bearing; Fracture
Citation
KNEE, v.21, no.1, pp.340 - 342
Indexed
SCIE
SCOPUS
Journal Title
KNEE
Volume
21
Number
1
Start Page
340
End Page
342
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/99662
DOI
10.1016/j.knee.2013.07.012
ISSN
0968-0160
Abstract
Meniscal bearing fracture is a rare complication of phase III Oxford unicompartmental knee replacement (UKR). We report a case of a meniscal bearing fracture that occurred 7 years after phase III Oxford medial UKR The meniscal bearing showed uneven delamination of the polyethylene in the thinnest articular surface and an impingement lesion. This lesion initiated a fatigue crack that propagated to cause failure of the meniscal bearing. This is the first report of a meniscal bearing fracture without a posterior marker wire. (C) 2013 Elsevier B.V. All rights reserved.