Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty A Retrieval Study Using 3-Dimensional Laser Scanning

Authors
Yun, Ho HyunShon, Won YongYoon, Jung RoYang, Jae-HyukLim, Deuk-Soo
Issue Date
9월-2012
Publisher
CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
Keywords
hip; total hip arthroplasty; wear; wear measurement; PowerPoint method
Citation
JOURNAL OF ARTHROPLASTY, v.27, no.8, pp.1530 - 1537
Indexed
SCIE
SCOPUS
Journal Title
JOURNAL OF ARTHROPLASTY
Volume
27
Number
8
Start Page
1530
End Page
1537
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/107660
DOI
10.1016/j.arth.2011.12.025
ISSN
0883-5403
Abstract
The purpose of this study was to evaluate the reliability of the Power Point (PP) (2007 Version; Microsoft, Redmond, Wash) method for measuring polyethylene liner wear after total hip arthroplasty. Seventeen retrieved polyethylene liners were included in this study. Wear volumes were calculated using the PP, the Dorr and Wan, and 3-dimensional (3D) laser scanning methods. Spearman correlation coefficients for wear volume results indicated strong correlations between the PP and 3D laser scanning methods (range, 0.89-0.93). On the other hand, Spearman correlation analysis revealed only moderate correlations between the Dorr and Wan and 3D laser scanning methods (range, 0.67-0.77). The PP method can be used to monitor linear wear after total hip arthroplasty and could serve as an alternative method when computerized methods are not available.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE