Comparison of Laser and Conventional Lancing Devices for Blood Glucose Measurement Conformance and Patient Satisfaction in Diabetes Mellitusopen accessComparison of Laser and Conventional Lancing Devices for Blood Glucose Measurement Conformance and Patient Satisfaction in Diabetes Mellitus

Other Titles
Comparison of Laser and Conventional Lancing Devices for Blood Glucose Measurement Conformance and Patient Satisfaction in Diabetes Mellitus
Authors
Kim Jung A박민정Eyun Song노은박소영이다영김재영유지희서지아최경묵백세현류혜진김난희
Issue Date
11월-2022
Publisher
대한당뇨병학회
Keywords
Blood glucose self-monitoring; Diabetes mellitus; Lasers; Pain
Citation
Diabetes and Metabolism Journal, v.46, no.6, pp.936 - 940
Indexed
SCIE
SCOPUS
KCI
Journal Title
Diabetes and Metabolism Journal
Volume
46
Number
6
Start Page
936
End Page
940
URI
https://scholar.korea.ac.kr/handle/2021.sw.korea/146546
DOI
10.4093/dmj.2021.0293
ISSN
2233-6079
Abstract
Self-monitoring of capillary blood glucose is important for controlling diabetes. Recently, a laser lancing device (LMT-1000) that can collect capillary blood without skin puncture was developed. We enrolled 150 patients with type 1 or 2 diabetes mellitus. Blood sampling was performed on the same finger on each hand using the LMT-1000 or a conventional lancet. The primary outcome was correlation between glucose values using the LMT-1000 and that using a lancet. And we compared the pain and satisfaction of the procedures. The capillary blood sampling success rates with the LMT-1000 and lancet were 99.3% and 100%, respectively. There was a positive correlation (r=0.974, <i>P</i><0.001) between mean blood glucose levels in the LMT-1000 (175.8±63.0 mg/dL) and conventional lancet samples (172.5±63.6 mg/dL). LMT-1000 reduced puncture pain by 75.0% and increased satisfaction by 80.0% compared to a lancet. We demonstrated considerable consistency in blood glucose measurements between samples from the LMT-1000 and a lancet, but improved satisfaction and clinically significant pain reduction were observed with the LMT-1000 compared to those with a lancet.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medical Science > 1. Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Kim, Nan Hee photo

Kim, Nan Hee
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE